As a guest user you are not logged in or recognized by your IP address. You have
access to the Front Matter, Abstracts, Author Index, Subject Index and the full
text of Open Access publications.
The demand for extended care for people suffering from heart failure is omnipresent. Wearables providing continuous heart rate measurement through optical sensors are of great interest due to their ease of use without the need for medical staff and their low cost. In this study, seven wearables were tested in fifteen measurement runs, with a duration of fourteen-hour each, and compared to a reference sensor. By calculating the Pearson correlation and the root mean square error, as well as the graphical representation by a Bland Altman plot, it was found that these wearables lack sufficient accuracy and may not be suitable for medical purposes.
This website uses cookies
We use cookies to provide you with the best possible experience. They also allow us to analyze user behavior in order to constantly improve the website for you. Info about the privacy policy of IOS Press.
This website uses cookies
We use cookies to provide you with the best possible experience. They also allow us to analyze user behavior in order to constantly improve the website for you. Info about the privacy policy of IOS Press.