As a guest user you are not logged in or recognized by your IP address. You have
access to the Front Matter, Abstracts, Author Index, Subject Index and the full
text of Open Access publications.
Endoleaks are one of the major concerns in the long-term follow-up of endovascular aneurysm repair treatment (EVAR). Therefore, periodic monitoring is required to detect eventual damages in an implanted stent-graft. A monitoring system for post EVAR procedure based on inductive-coupling which avoids the need to resorting to more complex biomedical imaging systems is presented here. Endoleaks are detected with capacitive pressure sensors placed in the stent-graft and monitored externally after the measure of the oscillation frequency provided by the LC circuit created by sensors and inductive coupling.