Manual accuracy in microsurgery is reduced by tremor and limited access. A surgical approach through the middle ear also puts delicate structures at risk, while the surgeon is often working at an unergonomic position. At this point a micromanipulator could have a positive influence. A system was developed to measure “working accuracy”, time and precision during manipulation in the middle ear. 10 ENT-surgeons simulated a perforation of the stapedial footplate on a modified 3D print of a human skull in a mock OR. Each trial was repeated more than 200 times aiming manually and using a micro-manipulator. Data of over 4000 measurements was tested and graphically processed. Work strain was evaluated with a questionnaire.
Accuracy for manual and micromanipulator perforation revealed a small difference. Learning curves showed a stronger decrease both in deviation and time when the micromanipulator was used. Also a lower work strain was apparent. The micromanipulator has the potential as an aiding device in ear surgery.